您好,欢迎访问一九零五行业门户网

可程式高低温箱标准中规定的产品结构及外观要求

可程式高低温箱标准中规定的产品结构及外观要求
the product structure and appearance may be required to program the gdw standards
经常浏览本站的朋友会发现,我们关于可程式高低温箱标准中规定的相关要求,之前一直都是描述的标准起源与术语定义,以及设备的一些外部条件和内容环境等,今天,我们一起来了解下标准中对于产品结构与外观的要求,也希望此篇文章能给正在对可程式高低温箱型的广大新老客户们有一些小小的帮助,标准的就是的:高低温试验箱,高低温交变试验箱,高低温交变湿热试验箱。
often visit the site to friends will find, we related requirements may be required to program the gdw standards, until is to define the origin and terminology standard description, equipment and some external conditions and environment, today, we come to know the requirements for product structure and appearance of the standard, also hope this article can give is to programmable high-low temperature box type of the vast number of old and new customers with a little help, the standard is the authoritative: high and low temperature test, high temperature test box, alternating high and low temperature test chamber.
1.首先可程式高低温箱内壁应使用耐热不易氧化和具有一定机械强度的材料制造。应无影响试验的污染源。
manufacturing 1 first test box wall should use the heat is not easy to be oxidized and a certain mechanical strength of materials. should not affect the pollution source test.
2.保温材料应能耐高温并具有阻燃性能。保温层应有足够的厚度,能保证试验箱外部已触及部位的温度在高温试验时不高于50℃,在低温试验及环境温度为15℃~35℃、相对湿度≤85%时不应有凝露现象。
2 insulation material should resist high temperature and flame retardant properties. insulation shall be of sufficient thickness, can ensure the test box external has hit parts of the temperature is not higher than 50 ℃ in high temperature test, tests at low temperatures and environmental temperature is 15 ℃ ~35 ℃, relative humidity below 85%, there should be no condensation phenomenon.
3.加热和制冷器件和冷量不应直接辐射在试验样品上。
3 heating and cooling device and cooling capacity should not direct radiation on test specimens.
4.工作室应设有观察窗和照明装置。
studio 4 should be provided with an observation window and lighting device.
5.箱门应密封良好,密封条应有良好的抗高温老化、耐低温硬化性能。
the 5 door should be sealed well, seal shall have good resistance to high-temperature, low temperature resistance and hardening property.
6.制冷系统不应有漏气、漏水、漏油缺陷。
6 refrigeration system should not leak, leakage, water leakage defects.
7.应有放置或悬挂试验样品的样品架。样品架应有足够的耐高温、低温性能。
the sample holder should be placed or hung 7 test samples. the sample frame should have enough resistance to high temperature, low temperature performance.
8.应设有引线孔。
8 shall be provided with a pin hole.
9.外观涂镀层应平整光滑、色泽均匀,不得有露底、起泡、起层或擦伤痕迹。
9 the appearance of coating should be smooth, uniform color, not to betray a confidence, foaming, layer or scratches.
附:可程式高低温箱安全和环境保护要求:
attached: safety and environmental protection requirements:
(1).接线端子对箱体金属外壳之间的绝缘电阻值应满足:冷态2mω以上,热态1mω以上(用500v,准确度为1.0级兆欧表测量),并能承受50hz交流电压1500v、施压时间5s耐电压试验。
(1). the terminal shall meet the insulation resistance between the metal shell of the value of more than 2m omega: cold, hot 1m ω above (with 500v, the accuracy is 1 grade megger measurement), and can withstand 50hz ac voltage 1500v, time 5s voltage withstanding test pressure
(2).保护接地端子应与试验箱外壳有良好的电气连接并能方便牢固的地接线,应符合gb/t14048.1—2000的7.1.9的规定。
(2). the protective earth terminal should have good electrical connection and convenient for firm connection and test case, shall be in accordance with the provisions of the 7.1.9 gb/t14048.1 2000.
(3).应有超温、过电流、缺水等保护及报警装置。
(3). and the alarm should be over temperature, over-current protection device, water etc..
(4).整机噪声应不高于80db(a)。
(4). the noise should be not more than 80db (a).